Introduction
Spaulding classification for medical devices and level of disinfection
- •Critical: A device that enters normally sterile tissue or the vascular system. Such devices should be sterilized, defined as the destruction of all microbial life. Examples of endoscopic instruments that require sterilization are biopsy forceps and papillotomes.
- •Semicritical: A device that comes into contact with intact mucous membranes and does not ordinarily penetrate sterile tissue. These devices (e.g., endoscopes) should receive at least high-level disinfection, defined as the destruction of all vegetative microorganisms, mycobacteria, small or nonlipid viruses, medium or lipid viruses, fungal spores, and some but not all bacterial spores.
- •Noncritical: Devices that do not ordinarily touch the patient or touch only intact skin, such as stethoscopes or patient carts. These items may be cleaned by low-level disinfection.
Pathogen transmission
GI Endoscope Reprocessing
Recommendations
- 1.All healthcare personnel in the endoscopy suite should be trained in and adhere to standard infection control recommendations (e.g., standard precautions), including those to protect both patients and healthcare workers. Category IA9
- 2.Perform pressure/leak testing after each use according to manufacturer guidelines. Category IB11,12,13,14,21
- 3.Disconnect and disassemble endoscope components (e.g., air/water and suction valves) as far as possible and completely immerse the endoscope and components in the enzymatic detergent. Category 1B21,22,23
- 4.Cleaning is essential before manual or automated disinfection. Meticulously clean the entire endoscope, including valves, channels, connectors, and all detachable parts with an enzymatic detergent compatible with the endoscope immediately after use, according to the manufacturer's instructions. Flush and brush all accessible channels to remove all organic (e.g., blood, tissue) and other residues. Repeatedly actuate the valves during cleaning to facilitate access to all surfaces. Clean the external surfaces and components of the endoscope with a soft cloth, sponge, or brushes. Category IA2,10,11,12,13,14,21,24,25,26,27,28,29,30,31,32,33
- 5.Use brushes appropriate for the size of the endoscope channel, parts, connectors, and orifices (e.g., bristles should contact all surfaces) for cleaning. Cleaning items should be disposable or thoroughly cleaned and disinfected/sterilized between uses. Category II14,21,28,33
- 6.Discard enzymatic detergents after each use because these products are not microbicidal and will not retard microbial growth. Category IB12,14,21,23
- 7.Reusable endoscopic accessories (e.g., biopsy forceps or other cutting instruments) that break the mucosal barrier should be mechanically cleaned as described above and then sterilized between each patient use (high-level disinfection is not appropriate). Category IA2,5,9,10,11,12,13,14,21,26,28,33,34
- 8.Ultrasonic cleaning of reusable endoscopic accessories and endoscope components may be used to remove soil and organic material from hard-to-clean areas. Category II21,33
- 9.Endoscopes (and accessories) that come in contact with mucous membranes are classified as semicritical items and should receive at least high-level disinfection after each patient use. Category IA2,9,10,11,12,14,21,26,28,33,35
- 10.Use a high-level disinfectant/sterilant cleared by the Food and Drug Administration (FDA) for high-level disinfection/sterilization (www.fda.gov/cdrh/ode/germlab.html ). Category IA2,9,10,11,12,14,21,26,28,35,36
- 11.The exposure time and temperature for disinfecting semicritical patient care equipment varies among the FDA-cleared high-level disinfectants. Follow the FDA-cleared label claim for high-level disinfection unless several well-designed experimental scientific studies, endorsed by professional societies, demonstrate an alternative exposure time is effective for disinfecting semicritical items. The FDA label claim for high-level disinfection with >2% glutaraldehyde at 25°C ranges from 20 to 90 minutes depending on the product. However, multiple scientific studies and professional organizations support the efficacy of >2% glutaraldehyde at 20 minutes at 20°C. Category IA1,10,21,26,27,32,37,38,39,40,41,42,43,44,45,46,47,48,49
- 12.Select a disinfectant/sterilant that is compatible with the endoscope. The use of specific high-level disinfectants/sterilants on an endoscope should be avoided if the endoscope manufacturer warns against use because of functional damage (with or without cosmetic damage). Category IB21,50,51
- 13.The selection and use of disinfectants in the healthcare field is dynamic, and products may become available that were not in existence when this guideline was written. As newer disinfectants become available, persons or committees responsible for selecting disinfectants for GI endoscope reprocessing should be guided by products cleared by the FDA and information in the scientific literature. Category II21,26,36
- 14.Completely immerse the endoscope and endoscope components in the high-level disinfectant/sterilant and ensure all channels are perfused. Nonimmersible GI endoscopes should be phased out immediately. Category IB10,11,13,14,21,26,28,52,53,54
- 15.If an automated endoscope washer-disinfector (AEWD) is used, ensure that the endoscope and endoscope components can be effectively reprocessed in the AEWD (e.g., the elevator wire channel of duodenoscopes is not effectively disinfected by most AEWDs, and this step must be performed manually). Users should obtain and review model-specific reprocessing protocols from both the endoscope and AEWD manufacturers and check for compatibility. Category IB10,12,14,21,28,52,53,54,55,56
- 16.If an AEWD is used, place the endoscope and endoscope components in the reprocessor and attach all channel connectors according to the AEWD and endoscope manufacturers' instructions to ensure exposure of all internal surfaces with the high-level disinfectant/chemical sterilant. Category IB14,21,52,53,54
- 17.If an AEWD cycle is interrupted, high-level disinfection or sterilization cannot be assured. Category II14
- 18.Because design flaws have compromised the effectiveness of AEWDs, the infection control staff should routinely review FDA advisories and the scientific literature for reports of AEWD deficiencies that may lead to infection. Category II53,57,58,59,60
- 19.After high-level disinfection, rinse the endoscope and flush the channels with sterile, filtered, or tap water to remove the disinfectant/sterilant. Discard the rinse water after each use/cycle. Flush the channels with 70% to 90% ethyl or isopropyl alcohol and dry with forced-air. The final drying steps greatly reduce the possibility of recontamination of the endoscope by waterborne microorganisms. Category IA10,11,12,13,21,26,30,58,61,62,63
- 20.When storing the endoscope, hang it in a vertical position to facilitate drying (with caps, valves, and other detachable components removed as per manufacturer instructions). Category II11,12,14,21,26,28,64
- 21.Endoscopes should be stored in a manner that will protect the endoscope from contamination. Category II11,14,21,26,28
- 22.High-level disinfect or sterilize the water bottle (used for cleaning the lens and irrigation during the procedure), and its connecting tube at least daily. Sterile water should be used to fill the water bottle. Category IB11,21,65,66,67,68
- 23.Maintain a log indicating for each procedure the patient's name and medical record number (if available), the procedure, the endoscopist, and the serial number or other identifier of the endoscope (and AEWD, if used) to assist in an outbreak investigation. Category II11,14,21
- 24.Perform routine testing of the liquid sterilant/high-level disinfectant to ensure minimal effective concentration (MEC) of the active ingredient. Check the solution at the beginning of each day of use (or more frequently) and document the results. If the chemical indicator shows that the concentration is less than the MEC, the solution should be discarded. Category IA10,11,12,14,21,26,35,42
- 25.Discard the liquid sterilant/high-level disinfectant at the end of its reuse life (which may be single-use) regardless of the MEC. If additional liquid sterilant/high-level disinfectant is added to an AEWD (or basin, if manually disinfected), the reuse life should be determined by the first use/activation of the original solution, that is, the practice of “topping off” of a liquid sterilant/high-level disinfectant pool does not extend the reuse life of the liquid sterilant/high-level disinfectant. Category IB14,26,69
- 26.Facilities where endoscopes are used and disinfected should be designed to provide a safe environment for healthcare workers and patients. Air-exchange equipment (e.g., ventilation system, exhaust hoods) should be used to minimize the exposure of all persons to potentially toxic vapors (e.g., glutaraldehyde). The vapor concentration of the chemical sterilant used should not exceed allowable limits (e.g., those of the American Conference of Governmental Industrial Hygienists, Occupational Safety and Health Administration [OSHA]). Although organic vapor respirators appropriate for chemical exposures can provide respiratory protection (e.g., in the event of spills), they are not intended for routine use and are not a substitute for adequate ventilation, vapor recovery systems, and work practice controls. Category IB, IC10,11,14,21,70, ,72,73
- 27.Personnel assigned to reprocess endoscopes should receive device-specific reprocessing instructions (i.e., endoscope and/or AEWD manufacturer, as needed) to ensure proper cleaning and high-level disinfection or sterilization. Competency testing of personnel reprocessing endoscopes should be done on a regular basis (e.g., commencement of employment, annually). Temporary personnel should not be allowed to reprocess endoscopes until competency has been established. Category IA10,11,12,14,21
- 28.All personnel who use chemicals should be educated about the biologic and chemical hazards present while performing procedures that use disinfectants. Category IC21,74
- 29.Personal protective equipment (e.g., gloves, gowns, eyewear, respiratory protection devices) should be readily available and should be used, as appropriate, to protect workers from exposure to chemicals, blood, or other potentially infectious material (OPIM). Category IC21,74,75,76
- 30.Healthcare facilities should develop protocols to ensure that users can readily identify whether an endoscope is contaminated or is ready for patient use. Category II
- 31.The utility of routine environmental microbiologic testing of endoscopes for quality assurance has not been established. No recommendation.21
- 32.If environmental microbiologic testing is performed, standard microbiologic techniques should be used. Category II21,77
- 33.In the setting of an outbreak caused by a suspected infectious or chemical etiology, the environmental sampling should be performed according to standard outbreak investigation. Category IA11,21,78
- 34.Endoscopy-related infections should be reported to:
- 1.Persons responsible for infection control at the institution.
- 2.The appropriate public health agency (state or local health department as required by state law or regulation).
- 3.The Food and Drug Administration (www.fda.gov/medwatch ).
- 4.The Centers for Disease Control and Prevention (CDC).
- 5.The manufacturer(s) of the endoscope, disinfectant/sterilant, and AEWD (if used) Category IB, IC10,11,21,79
- FDA
Medical device reporting (MDR).http://www.fda.gov/cdrh/mdr/Date: November 20, 2002
Summary
Consensus Panel Members and Conference Participants
Disclaimer
Appendix
The CDC system for categorizing recommendations is as follows: |
Category IA: Strongly recommended for implementation and strongly supported by well-designed experimental, clinical, or epidemiologic studies. |
Category IB: Strongly recommended for implementation and supported by some experimental, clinical, or epidemiologic studies and a strong theoretical rationale. |
Category IC: Required by state or federal regulations. Because of state differences, readers should not assume that the absence of an IC recommendation implies the absence of state regulations. |
Category II: Recommended for implementation and supported by suggestive clinical or epidemiologic studies or theoretical rationale. |
No recommendation : Unresolved issue. Practices for which insufficient evidence or no consensus regarding efficacy exists. |
References
- Disinfection of medical and surgical materials.in: Disinfection, sterilization, and preservation. Lippincott Williams & Wilkins, Philadelphia2001: 881-917
- Transmission of infection by gastrointestinal endoscopy and bronchoscopy.Ann Intern Med. 1993; 118: 117-128
- Transmission of infection by gastrointestinal endoscopy.Gastrointest Endosc. 1993; 36: 885-888
- Transmission of Trichosporon asahii esophagitis by a contaminated endoscope.Mycoses. 2001; 44: 13-21
- Patient-to-patient transmission of hepatitis C virus during colonoscopy.N Engl J Med. 1997; 337: 237-240
- Nosocomial transmission of hepatitis C virus [letter].Ann Intern Med. 1999; 131: 794
- Hépatite C après cholangiographie rétrograde [French].Gastroenterol Clin Biol. 1993; 17: 763-775
- Draft guidance for the content of premarket notifications for endoscopes used in gastroenterology and urology.FDA, Rockville1995
- CDC guideline for handwashing and hospital environmental control, 1985.Infect Control Hosp Epidemiol. 1986; 7: 231-243
- Reprocessing of flexible gastrointestinal endoscopes.Gastrointest Endosc. 1996; 43: 540-546
- APIC guidelines for infection prevention and control in flexible endoscopy.Am J Infect Control. 2000; 28: 138-155
- Standard practice for cleaning and disinfection of flexible fiberoptic and video endoscopes used in the examination of the hollow viscera.West Conshohocken. 2000; : F1518-F1600
- Recommended practices for use and care of endoscopes. 2002 Standards, recommended practices, and guidelines. AORN, Denver2002: 229-232
- Standards of infection control in reprocessing of flexible gastrointestinal endoscopes.Gastroenterol Nurs. 2000; 23: 172-187
- Infection control practices in gastrointestinal endoscopy in the United States: a national survey.Infect Control Hosp Epidemiol. 1991; 12: 289-296
- Multi-state investigation of the actual disinfection/sterilization of endoscopes in health care facilities.Am J Med. 1992; 92: 257-261
- Variability in reprocessing policies and procedures for flexible fiberoptic endoscopes in Massachusetts hospitals.Am J Infect Control. 1992; 20: 283-290
- Correction of deficiencies in flexible fiberoptic sigmoidoscope cleaning and disinfection technique in family practice and internal medicine offices.Arch Fam Med. 1997; 6: 578-582
- GI endoscopic reprocessing practices in the United States.Gastrointest Endosc. 1999; 50: 362-368
- Current instrument reprocessing practices: results of a national survey.Gastroenterol Nurs. 2001; 24: 253-260
Rutala WA, Weber DJ, and the Healthcare Infection Control Practices Advisory Committee. Guideline for disinfection and sterilization in healthcare facilities. Am J Infect Control. In Press.
- Safety and cleaning of medical materials and devices.J Biomed Mater Res. 2000; 53: 131-136
- Endoscope reprocessing: where do we go from here?.J Hosp Infect. 1995; 30: 543-551
- Contamination of endoscopes used in AIDS patients.Lancet. 1989; 2: 86-88
- Signficant factors in the disinfection and sterilization of flexible endoscopes.Am J Infect Control. 1992; 20: 291-300
- APIC guideline for selection and use of disinfectants.Am J Infect Control. 1996; 24: 313-342
- Mycobacteria and glutaraldehye: is high-level disinfection of endoscopes possible?.Gastrointest Endosc. 1996; 43: 451-456
- Cleaning and disinfection of equipment for gastrointestinal endoscopy. Report of a Working party of the British Society of Gastroenterology Endoscopy Committee.Gut. 1998; 42: 585-593
- Natural bioburden levels detected on flexible gastrointestinal endoscopes after clinical use and manual cleaning.Gastrointest Endosc. 1998; 48: 137-142
- Antimicrobial efficacy of endoscopic disinfection procedures: a controlled, multifactorial investigation.Gastrointest Endosc. 1999; 50: 152-158
- High-level disinfection of gastrointestinal endoscopes: are current guidelines adequate?.Am J Gastroenterol. 1999; 94: 1546-1550
- Microbial bioburden in endoscope reprocessing and an in-use evaluation of the high-level disinfection capabilities of Cidex PA.Gastroenterology Nursing. 1999; 22: 63-68
- Guidelines on cleaning and disinfection in GI endoscopy.Endoscopy. 2000; 32: 77-83
- Disinfection of biopsy forceps and culture of Helicobacter pylori from gastric mucosal biopsies [letter].Am J Gastroenterol. 1999; 94: 1422-1423
- Recommended practices for high-level disinfection. 2002 Standards, recommended practices, and guideline. AORN, Denver2002: 211-216
FDA. FDA-cleared sterilants and high level disinfectants with general claims for processing reusable medical and dental devices 1/30/02. Available at: http://www.fda.gov/cdrh/ode/germlab.html. Accessed 12-9-02.
- Kinetics of tuberculocidal response by alkaline glutaraldehye in solution and on an inert surface.J Appl Bacteriol. 1986; 61: 87-93
- Bactericidal activity of alkaline glutaraldehyde solution against a number of atypical mycobacterial species.J Appl Bacteriol. 1986; 61: 247-251
- A more accurate method for measurement of tuberculocidal activity of disinfectants.Appl Environ Microbiol. 1987; 53: 2189-2192
- Use of membrane filters for measurement of mycobacterial activity of alkaline glutaraldehyde solution.Appl Environ Microbiol. 1987; 53: 737-739
- Efficacies of selected disinfectants against Mycobacterium tuberculosis.J Clin Microbiol. 1990; 28: 2234-2239
- Effect of methodology, dilution, and exposure time on the tuberculocidal activity of glutaraldehyde-based disinfectants.Appl Environ Microbiol. 1990; 56: 1813-1817
- Elimination of high titre HIV from fiberoptic endoscopes.Gut. 1990; 31: 657-659
- Viral transmission and fibreoptic endoscopy.J Hosp Infect. 1991; 18: 136-140
- Inactivation of Mycobacterium tuberculosis and Mycobacterium bovis by 14 hospital disinfectants.Am J Med. 1991; 91: 267S-271S
- A study of glutaraldehyde disinfection of fibreoptic bronchoscopes experimentally contaminated with Mycobacterium tuberculosis.J Hosp Infect. 1992; 22: 137-142
- Feasibility of a combined carrier test for disinfectants: studies with a mixture of five types of microorganisms.Am J Infect Control. 1994; 22: 152-162
- Mycobacterium gordonae in fiberoptic bronchoscopes.Am J Infect Control. 1996; 24: 19-23
- Effectiveness of a manual disinfection procedure in eliminating hepatitis C virus from experimentally contaminated endoscopes.Gastrointest Endosc. 1999; 50: 147-151
- Liquid sterilization versus high level disinfection in the urologic office.Urology. 1997; 50: 337-340
- Disinfection of endoscopes: review of new chemical sterilants used for high-level disinfection.Infect Control Hosp Epidemiol. 1999; 20: 69-76
- Importance of lumen flow in liquid chemical sterilization.Am J Infect Control. 1999; 20: 458-459
- Nosocomial tranmission of imipenem-resistant Pseudomonas aeruginosa following bronchoscopy associated with improper connection to the STERIS System 1 processor.Infect Control Hosp Epidemiol. 2001; 20: 514-516
- Bronchoscopy-related infections and pseudoinfections—New York, 1996 and 1998.MMWR. 1999; 48: 557-560
- Pseudomonas aeruginosa infection during endoscopy: reply [letter].Gastroenterology. 1987; 93: 1451
- Pseudomonas aeruginosa and Enterobacteriaceae bacteremia after biliary endoscopy: an outbreak investigation using DNA macrorestriction analysis.Am J Med. 1993; 95: 489-498
- Pseudomonas aeruginosa infection during endoscopy [letter].Gastroenterology. 1987; 93: 1451
- Nosocomial infections from contaminated endoscopes: a flawed automated endoscope washer. An investigation using molecular epidemiology.Am J Med. 1991; 91: 272S-280S
- Nosocomial infection and pseudoinfection from contaminated endoscopes and bronchoscopes—Wisconsin and Missouri.MMWR. 1991; 40: 675-678
- Contamination of flexible fiberoptic bronchoscopes with Mycobacterium chelonae linked to an automated bronchoscope disinfection machine.Am Rev Respir Dis. 1992; 145: 853-855
- Cleaning and disinfection of fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and forced-air drying.Gastroenterology. 1982; 83: 613-618
- Pseudomonas infection of the biliary system resulting from use of a contaminated endoscope.Gastroenterology. 1987; 92: 759-763
- In hospital evaluation of contamination of duodenoscopes: a quantitative assessment of the effects of drying.J Hosp Infect. 1991; 19: 89-98
- The significance of bacterial contamination of fiberoptic endoscopes.J Hosp Infect. 1980; 1: 53-61
- Water-related nosocomial pneumonia caused by Legionella pneumophila serogroups 1 and 10.J Infect Dis. 1985; 152: 356-364
- Mycobacterium xenopi and Mycobacterium kansasii in a hospital water supply.J Hosp Infect. 1985; 6: 175-178
- Water as a reservoir of nosocomial pathogens.Infect Control Hosp Epidemiol. 1997; 18: 609-616
- Nosocomial outbreaks/pseudo-outbreaks caused by nontuberculous mycobacteria.Annu Rev Microbiol. 1998; 52: 453-490
- Transmission of infection by gastrointestinal endoscopy.Gastrointest Endosc. 2001; 54: 824-828
- Expanding role of hospital epidemiology: employee health—chemical exposure in the health care setting.Infect Control Hosp Epidemiol. 1989; 10: 261-266
- Air contaminants final rule.Federal Register. 1993; 58: 35338-35351
- Occupational risks associated with the use of selected disinfectants and sterilants.in: Disinfection, sterilization, and antisepsis in healthcare. Polyscience Publications, Champlain1998: 211-226
- Threshold limit values for chemical substances and physical agents and biologic exposure indices. Cincinnati, 2001
OSHA. Hazard Communication Standard.29 CFR 1910.1200. Washington, D.C.
- Occupational exposure to bloodborne pathogens: final rule.Federal Register. 1991; 56: 64003-64182
- Per msonal protective equipment.Gastrointest Endosc. 1999; 49: 854-857
- Microbiological culturing of environment and medical-device surfaces.in: Clinical microbiology procedures handbook. American Society for Microbiology, Washington (DC)1992: 11.0.1-1109
- Investigation of endemic and epidemic nosocomial infection data.in: 3rd ed. Hospital infection. Little Brown, Boston1992: 109-135
- Medical device reporting (MDR).(Available at)http://www.fda.gov/cdrh/mdr/(Accessed)Date: November 20, 2002
Article Info
Footnotes
*Reprinted from Gastrointest Endosc (2003;58:1-8) and Infect Control Hosp Epidemiol (2003;27:532-7).
**Am J Infect Control 2003;31:309-15.