Relationship between lumen diameter and length sterilized in the 125L ozone sterilizer

Published:January 31, 2008DOI:https://doi.org/10.1016/j.ajic.2007.06.006

      Background

      A safe alternative to ethylene oxide for reprocessing heat-sensitive lumen medical devices, such as endoscopes, is needed. The effectiveness of a new, safe, low-cost, and environmentally friendly low-temperature sterilization process using ozone was assessed.

      Method

      Rigid lumen devices were used to assess the maximum length of lumens of different internal diameters that can be sterilized in the TSO3 model 125L ozone sterilizer. Two inoculation techniques were used. An inoculated wire was placed inside lumens with internal diameters of 0.8 mm and larger, whereas lumens with an internal diameter of 0.5 mm were inoculated directly.

      Results

      Lumens with internal diameters of 0.5 mm, 1 mm, 2 mm, 3 mm, and 4 mm with lengths varying between 45 and 70 cm can be sterilized with ozone. Calculation of the log reduction value for each size demonstrated the achievement of a sterility assurance level of 10−6. Experimental results demonstrated a linear relationship (with r2 = 0.990) between the length of lumen that can be sterilized in the 125L ozone sterilizer and its internal diameter. Effective sterilization of an ACMI ureteroscope that is more challenging in terms of sterilant penetration in a small lumen (0.8 mm) compared with the stated lumen claims confirms that the relationship can conservatively be used to predict the length of a lumen device that can be sterilized in the 125L ozone sterilizer for a given diameter.

      Conclusion

      Intermediate sizes of lumen devices that can be sterilized in the 125L ozone sterilizer can be interpolated from the linear relationship between diameter and length found in the present study.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to American Journal of Infection Control
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Centers for Disease Control and Prevention
        Guidelines for handwashing and hospital environmental control. Section 2: Cleaning, disinfecting, and sterilizing patient care equipment.
        US Department of Health and Human Services, Atlanta (GA)1985
      1. US Pharmacopeia. 〈1211〉: Sterilization and quality assurance. In: US Pharmacopeia National Formulary (USP29/NF24).
        United States Pharmacopeial Convention, Rockville (MD)2006 (p. 2811–14)
      2. US Pharmacopeia. 〈1035〉: Biological indicators for sterilization. In: U.S. Pharmacopeia National Formulary (USP29/NF24).
        United States Pharmacopeial Convention, Rockville (MD)2006 (p. 2811–14)
      3. ANSI/AAMI/ISO 14161: Sterilization of health care products: biological indicators. Guidance for the selection, use, and interpretation of results.
        Association for the Advancement of Medical Instrumentation, Arlington (VA)2000 (p. 1–49)
      4. ANSI/AAMI/ISO 11737-1: Sterilization of a health care products: microbiological methods. Part 1: determination of the population of microorganisms on product.
        Association for the Advancement of Medical Instrumentation, Arlington (VA)2006 (p. 1–29)
        • Gillis J.R.
        • Schmidt W.C.
        Scanning electron microscopy of spores on inoculated product surfaces.
        Med Device Diagn Indust. 1993; 5: 46-49
        • Environment Canada
        Guidelines for the reduction of ethylene oxide releases from sterilization applications.
        Canadian Department of the Environment, Ottawa, Canada2005 (p. 1-9)
        • Occupational Safety and Health Administration
        Occupational exposure to ethylene oxide. Final standard 29CFR1910.
        Fed Regist. 1984; 49: 25734
      5. ISO 10993-7: Biological evaluation of medical devices. Part 7: Ethylene oxide sterilization residuals.
        Switzerland: International Organization for Standardization, Geneva1995 (p. 1–40)
      6. Advanced Sterilization Products. 100S domestic lumens chart. Available at http://www.sterrad.com/Products_&_Services/STERRAD/Files/100S_Domestic_Lumen.pdf. Accessed June 12, 2007.

      7. Advanced Sterilization Products. Frequently asked question: What devices and materials can be sterilized in the STERRAD NX System? Available at: http://www.sterrad.com/Products_&_Services/STERRAD/STERRAD_NX/FAQs/. Accessed June 12, 2007.

        • Chaunet M.
        • Dufresne S.
        • Robitaille S.
        Sterilization technology for the 21st century.
        TSO3 Inc, Québec, Canada2007 (p. 1–21)
        • Pflug I.J.
        Microbiology and engineering of sterilization processes.
        10th ed. University of Minnesota, Department of Food Science and Nutrition, Environmental Sterilization Laboratory, Minneapolis (MN)1999 (p. 1.1-22.9)
      8. ISO 11135-1: Sterilization of health care products: ethylene oxide. Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
        Switzerland: International Organization for Standardization, Geneva2007 (p. 1–41)
      9. ANSI/AAMI/ISO 17665-1: Sterilization of health care products: moist heat. Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
        Association for the Advancement of Medical Instrumentation, Arlington (VA)2006 (p. 1–43)
      10. ANSI/AAMI/ISO 11138-1: Sterilization of health care products: biological indicators. Part 1: General.
        Association for the Advancement of Medical Instrumentation, Arlington (VA)2006 (p. 1–43)