Antiseptic barrier caps to prevent central line-associated bloodstream infections: a systematic review and meta-analysis

Open AccessPublished:September 15, 2022DOI:https://doi.org/10.1016/j.ajic.2022.09.005

      Highlights

      • A safe and reliable central venous access is the cornerstone of treatments in numerous clinical settings.
      • Antiseptic barrier caps seems an effective strategy to reduce CLABSI
      • Antiseptic barrier caps are safe, timesaving and highly appreciated by health care workers for their ease of use.
      • Monitoring of ABC use, and availability of caps is key to improve and maintain compliance.

      Abstract

      Background

      Reliable and safe venous access is crucial for patients using central venous catheters (CVC). However, such CVCs carry a risk for central line-associated bloodstream infections (CLABSIs). Antiseptic barrier caps (ABCs) are a novel tool in the armamentarium for CVC disinfection. Our aim was to review the efficacy and safety of ABCs.

      Method

      A literature search was conducted using PubMed, EMBASE, Cochrane library, and CINAHL. Primary aim was to compare CLABSI rates in patients using ABCs versus standard care. Secondary aims included efficacy of ABCs in relevant subgroups (age, ABC brand, clinical setting), safety, compliance, and costs. Fifteen studies were included in the meta-analysis.

      Results

      In total, 391 CLABSIs in 273,993 catheter days occurred in the intervention group versus 620 CLABSIs in 284,912 days in the standard care group, resulting in a risk ratio of 0.65 (95%CI 0.55-0.76; P<0.00001). Subgroup analyses showed similar effects, except for non-intensive care unit. In general, ABCs were safe, highly appreciated by patients and caregivers, and cost-effective, while compliance was easy to monitor. In most studies, a substantial risk of bias was observed.

      Conclusion

      In conclusion, while available evidence suggests that ABCs are effective, safe, easy in use, and cost-effective. However, due to the poor methodological quality of most available studies, more robust data should justify their use at this point.

      Graphical abstract

      Keywords

      Introduction

      A reliable and safe vascular access by means of central venous catheters (CVCs) is crucial for the care of patients in numerous clinical settings, ranging from infusion of parenteral nutrition and medication to haemodialysis.
      • Maki DG
      • Kluger DM
      • Crnich CJ.
      The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies.
      ,
      • Ruesch S
      • Walder B
      • Tramèr MR.
      Complications of central venous catheters: internal jugular versus subclavian access–a systematic review.
      Despite their vital importance, these CVCs may cause potentially life-threatening complications, mostly central line-associated bloodstream infections (CLABSIs).
      • Pronovost P
      • Needham D
      • Berenholtz S
      • Sinopoli D
      • Chu H
      • Cosgrove S
      • et al.
      An intervention to decrease catheter-related bloodstream infections in the ICU.
      ,
      • Ziegler MJ
      • Pellegrini DC
      • Safdar N.
      Attributable mortality of central line associated bloodstream infection: systematic review and meta-analysis.
      In case of a CLABSI, pathogens may enter the circulation via an extraluminal or the intraluminal route (Figure 1). Contamination via the extraluminal route is most common for short-term (non-tunneled) CVCs when micro-organisms migrate from the skin at the insertion site directly into the vein. To prevent extraluminal contamination, the skin is disinfected prior to catheter insertion.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      Contamination via the intraluminal route is the most frequent complication seen in long-term subcutaneously tunneled CVCs and frequently occurs due to inadequate antiseptic catheter-handling procedures.
      • de Cicco M
      • Panarello G
      • Chiaradia V
      • Fracasso A
      • Veronesi A
      • Testa V
      • et al.
      Source and route of microbial colonisation of parenteral nutrition catheters.
      ,
      • Frasca D
      • Dahyot-Fizelier C
      • Mimoz O.
      Prevention of central venous catheter-related infection in the intensive care unit.
      The intraluminal route starts at the catheter hub or (if present) needleless connector, from which micro-organisms can adhere, migrate, and colonize the internal lumen of a CVC. These micro-organisms may form a biofilm and subsequently spread into the bloodstream.
      • Frasca D
      • Dahyot-Fizelier C
      • Mimoz O.
      Prevention of central venous catheter-related infection in the intensive care unit.
      • Bond A
      • Chadwick P
      • Smith TR
      • Nightingale JMD
      • Lal S.
      Diagnosis and management of catheter-related bloodstream infections in patients on home parenteral nutrition.
      • Mermel LA
      • Allon M
      • Bouza E
      • Craven DE
      • Flynn P
      • O'Grady NP
      • et al.
      Clinical practice guidelines for the diagnosis and management of intravascular catheter-related infection: 2009 Update by the Infectious Diseases Society of America.
      To prevent intraluminal contamination, catheter hubs are disinfected according to standard manual procedures, called “scrub-the-hub”. The hub is scrubbed with a dis-infectant wipe and subsequently dried before using the catheter. Since scrubbing duration is generally not described in protocols and compliance differs among healthcare professionals, inadequate disinfection results in an increased risk of CLABSIs.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Casey AL
      • Karpanen TJ
      • Nightingale P
      • Elliott TSJ.
      An in vitro comparison of standard cleaning to a continuous passive disinfection cap for the decontamination of needle-free connectors.
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.

      Vries d. Reducing Bloodstream Infection Risk in Central and Peripheral Intravenous Lines: Initial Data on Passive Intravenous Connector Disinfection. 2014.

      Figure 1
      Figure 1Central venous catheter with antiseptic barrier cap
      Antiseptic barrier caps (ABCs) have been developed to decrease CLABSIs by reducing the effect of variations in scrubbing duration and techniques. These plastic caps contain a disinfectant, mostly 70% isopropyl alcohol (IPA) or chlorhexidine gluconate. The cap is screwed directly onto the needle-less connector and continuously bathes the access point in an antimicrobial agent and can be left in place between infusions, providing improved disinfection and preventing touch or airborne pathogens from invading the hub (Figure 1).
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      -
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      As long as the cap is not removed from the access point, the needle-less connector remains inaccessible, disinfected and protected for up to seven days.
      • Inchingolo R
      • Pasciuto G
      • Magnini D
      • Cavalletti M
      • Scoppettuolo G
      • Montemurro G
      • et al.
      Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit 11 Medical and Health Sciences 1117 Public Health and Health Services 11 Medical and Health Sciences 1103 Clinical Sciences.
      -
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      Most protocols recommend ABCs to be replaced each time the CVC is accessed, or at least once weekly. Hence, ABCs seem a less error-prone approach to prevent CLABSIs as a result of their single-use. The time-saving feature is highly appreciated by healthcare professionals, resulting in high compliance with cap use.
      • Tejada S
      • Leal-Dos-Santos M
      • Peña-López Y
      • Blot S
      • Alp E
      • Rello J.
      Antiseptic barrier caps in central line-associated bloodstream infections: A systematic review and meta-analysis.
      Implementation of these caps could provide a considerable help to prevent CLABSIs in clinical practice, as was described in a relative recent meta-analysis.
      • Tejada S
      • Leal-Dos-Santos M
      • Peña-López Y
      • Blot S
      • Alp E
      • Rello J.
      Antiseptic barrier caps in central line-associated bloodstream infections: A systematic review and meta-analysis.
      As several studies with the highest level of evidence on ABCs have been published even more recently, an updated review and meta-analysis seems appropriate. Primary aim of this study was to assess the efficacy and safety of ABCs to prevent CLABSIs. Secondary aims were to evaluate efficacy in different clinical settings, safety, compliance, and costs.

      Methods

      This study was registered in PROSPERO (registration number: CRD42021288689) and reports according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines.

      Search strategy

      A first literature search was conducted on the 24th of November 2021, using MedLine, EMBASE, Cochrane library and CINAHL. The search was updated on 1 June 2022 to avoid missing recently published trials. The following search terms (and their synonyms) were combined: “barrier cap” and “central venous catheter”. The full search strategy is presented in the Table S1. An experienced medical librarian of the Radboud University assisted with the literature search.

      Study selection

      We included studies reporting prospective data on ABCs for CVCs, with CLABSI rates as primary or secondary outcome. Patients from any age group, in any clinical setting were eligible. Publications were only included in the meta-analysis if they reported at least two of the following variables: total catheter days, number of CLABSIs, and/or CLABSIs per 1,000 catheter days. Studies implementing a set of interventions to reduce CLABSI rates were only included if the implementation of ABCs was the sole intervention in a certain time period.
      Studies exclusively focusing on catheter connectors or lock solutions were excluded from this review, as well as articles that could not be retrieved, reviews, conference abstracts, letters to editors, case reports, corrigenda and product audits. ABC brands that are currently not commercially available (anymore) were excluded as well.

      Data extraction

      All identified studies were checked for duplicates, and title and abstract were screened. Included studies were subsequently screened for full text. Literature screening and data collection was independently performed by two authors (MvE and VG) and discrepancies were discussed with YW and GW. The following data was extracted from each article: publication date, study design, country, department, population (children and/or adults), study period, number of patients, number of CVC days, type of CVC, type of brand cap, CLABSI or catheter-related bloodstream infection (CRBSI) rates. In case of missing data, the corresponding or final author of a study was requested via email for additional information.

      Outcomes and definitions

      Primary outcome of this study was the rate of CLABSIs per 1,000 catheter days in patients using ABCs compared to standard (non-ABC) care, which includes chlorhexidine and IPA wipes. Secondary outcomes were the efficacy of ABCs in various settings, such as adults and children, type of cap brands, and (non-)intensive care unit (ICU) setting. Furthermore, we explored safety, compliance, and costs of the ABCs. Publications with the outcome CLABSI or CRBSI were both included and for readability purposes in this article referred to as CLABSI.

      Quality assessment of bias

      All randomized and non-randomized studies were assessed using the Risk-of-Bias for randomized trials tool (RoB 2) and the Risk Of Bias In Non-randomized Studies- of Interventions tool (ROBINS-I), respectively. Non-randomized studies were scored as low, moderate, serious, or critical risk of bias, and randomized studies as low, some concerns, and high risk of bias.
      • Sterne JA
      • Hernán MA
      • Reeves BC
      • Savović J
      • Berkman ND
      • Viswanathan M
      • et al.
      ROBINS-I: a tool for assessing risk of bias in non-randomised studies of interventions.
      • Sterne JAC
      • Savović J
      • Page MJ
      • Elbers RG
      • Blencowe NS
      • Boutron I
      • et al.
      RoB 2: a revised tool for assessing risk of bias in randomised trials.
      • McGuinness LA
      • Higgins JPT.
      Risk-of-bias VISualization (robvis): An R package and Shiny web app for visualizing risk-of-bias assessments.
      Both assessment were reviewed by two independent reviewers (VG and MvE)

      Statistical analysis

      The data of included studies in the meta-analyses were pooled and analyzed using the Mantel-Haenszel method with a random effect model. The main outcome of the meta-analysis was expressed in risk ratio (RR), based on catheter days and number of CLABSIs. Heterogenicity between studies was established with Higgins’ I2 statistics. A score between 0-40%, 30-60%, 50-90% and 75-100% was defined as might not be important, moderate, substantial and considerable, respectively.
      • Higgins JP
      • Thompson SG
      • Deeks JJ
      • Altman DG.
      Measuring inconsistency in meta-analyses.
      A funnel plot was used to determine publication bias, and sensitivity analyses were performed to establish potential factors influencing outcomes, such as type or quality of a study. In addition, subgroup analyses were performed for age (children versus adults), type of brand (Curos™ 3M, St. Paul, MN versus SwabCap™ Excelsior Medical, Neptune City, NJ), and patient population (ICU versus non-ICU), and tested for different effects between subgroups.
      • Borenstein M
      • Higgins JP.
      Meta-analysis and subgroups.
      All analyses were conducted using Review manager version 5.3.5.

      Results

      Search strategy

      Our search yielded 3,599 articles. After removing duplicates, 2,713 potentially relevant articles were screened for title and abstract. Subsequently, 42 articles were assessed for full text, of which 26 studies were excluded for various reasons (Figure 2). Of the remaining 16 studies, corresponding authors of ten articles were approached for additional data on outcomes or for the baseline table, of which two responded.
      • Inchingolo R
      • Pasciuto G
      • Magnini D
      • Cavalletti M
      • Scoppettuolo G
      • Montemurro G
      • et al.
      Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit 11 Medical and Health Sciences 1117 Public Health and Health Services 11 Medical and Health Sciences 1103 Clinical Sciences.
      ,
      • Tasdelen Ogulmen D
      • Ates S.
      Use of alcohol containing caps for preventing bloodstream infections: A randomized controlled trial.
      In addition, we approached authors of a previous meta-analysis, of Voor in ‘t Holt et al, and retrieved missing information from five studies.
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Kamboj M
      • Blair R
      • Bell N
      • Son C
      • Huang YT
      • Dowling M
      • et al.
      Use of Disinfection Cap to Reduce Central-Line-Associated Bloodstream Infection and Blood Culture Contamination Among Hematology-Oncology Patients.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      ,
      • Voor In 't Holt AF
      • Helder OK
      • Vos MC
      • Schafthuizen L
      • Sülz S
      • van den Hoogen A
      • et al.
      Antiseptic barrier cap effective in reducing central line-associated bloodstream infections: A systematic review and meta-analysis.
      • Sweet MA
      • Cumpston A
      • Briggs F
      • Craig M
      • Hamadani M
      Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line–associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit.
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      We were unable to retrieve outcome data from one remaining study.
      • Pavia M
      • Mazza M
      Adding innovative practices and technology to central line bundle reduces bloodstream infection rate in challenging pediatric population.
      Finally, 16 articles were reviewed systematically, of which 15 were included in the meta-analysis (Table 1 and Figure 2).
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      • Inchingolo R
      • Pasciuto G
      • Magnini D
      • Cavalletti M
      • Scoppettuolo G
      • Montemurro G
      • et al.
      Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit 11 Medical and Health Sciences 1117 Public Health and Health Services 11 Medical and Health Sciences 1103 Clinical Sciences.
      • Kamboj M
      • Blair R
      • Bell N
      • Son C
      • Huang YT
      • Dowling M
      • et al.
      Use of Disinfection Cap to Reduce Central-Line-Associated Bloodstream Infection and Blood Culture Contamination Among Hematology-Oncology Patients.
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      ,
      • Tasdelen Ogulmen D
      • Ates S.
      Use of alcohol containing caps for preventing bloodstream infections: A randomized controlled trial.
      ,
      • Sweet MA
      • Cumpston A
      • Briggs F
      • Craig M
      • Hamadani M
      Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line–associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit.
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      • Castello FV
      • Maher A
      • Cable G.
      Reducing bloodstream infections in pediatric rehabilitation patients receiving parenteral nutrition.
      • Cruz-Aguilar R
      • Carney J
      • Mondaini V
      • Vehreschild M
      • Griskaitis M
      • Salmanton-García J
      • et al.
      A quality improvement study on the reduction of central venous catheter-associated bloodstream infections by use of self-disinfecting venous access caps (STERILE).
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      Figure 2
      Figure 2Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) Flowchart
      • Page MJ
      • McKenzie JE
      • Bossuyt PM
      • Boutron I
      • Hoffmann TC
      • Mulrow CD
      • et al.
      The PRISMA 2020 statement: an updated guideline for reporting systematic reviews.
      Table 1Characteristics of 16 included studies
      StudyStudy designDepartment (country)PopulationStudy periodNumber of patientsNumber of line daysType of cathetersType of interventionOutcomeBSI rate
      CLABSI or CRBSI
      Risk of bias
      ROB2 or ROBINS-II
      Cameron-Watson et al. (2016)
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      Pre -post intervention designICU and non-ICU (United Kingdom)AdultsSC: 6 months I: 6 monthsSC: - I: -SC: 6,046 I: 5,333CVC, PICC and arterial VAD, PIVCuros capCRBSISC: 4.3 I: 1.5Critical
      Castello et al. (2011)
      • Castello FV
      • Maher A
      • Cable G.
      Reducing bloodstream infections in pediatric rehabilitation patients receiving parenteral nutrition.
      Pre -post intervention designNon-ICU (United states)ChildrenSC: ≈17 months I: ≈17 monthsSC: 5 I: 10SC: 416 I: 392CVCSwabCapCLABSISC: 4.8 I: 0.0Serious
      Cruz (2021)
      • Cruz-Aguilar R
      • Carney J
      • Mondaini V
      • Vehreschild M
      • Griskaitis M
      • Salmanton-García J
      • et al.
      A quality improvement study on the reduction of central venous catheter-associated bloodstream infections by use of self-disinfecting venous access caps (STERILE).
      Pre -post intervention designICU and non-ICU (Germany)AdultsSC: 12 months I: 12 monthsSC: 443 I: 431SC: 4,189 I: 4,818CVCCuros capCLABSISC: 15.28 I: 10.38Moderate
      Helder et al. (2020)
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      Pre -post intervention designICU (Netherlands)ChildrenSC: 24 months I: 12 monthsSC: 1,482 I: 766SC: 15,225 I: 7,366CVCCuros capCLABSISC: 3.2 I: 2.4Moderate
      Inchingolo et al. (2019)
      • Inchingolo R
      • Pasciuto G
      • Magnini D
      • Cavalletti M
      • Scoppettuolo G
      • Montemurro G
      • et al.
      Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit 11 Medical and Health Sciences 1117 Public Health and Health Services 11 Medical and Health Sciences 1103 Clinical Sciences.
      Pre -post intervention designICU (Italie)AdultsSC: 9 months I: 9 monthsSC: 86 I: 21SC: 1,041 I: 326CVCCuros capCLABSISC: 8.6 I: 0.0Moderate
      Kamboj et al. (2015)
      • Kamboj M
      • Blair R
      • Bell N
      • Son C
      • Huang YT
      • Dowling M
      • et al.
      Use of Disinfection Cap to Reduce Central-Line-Associated Bloodstream Infection and Blood Culture Contamination Among Hematology-Oncology Patients.
      Pre -post intervention designICU and non-ICU (United states)AdultsSC: 16 months I: 16 monthsSC: - I: -SC: 84,427 I: 83,659CVCSwabCapCLABSISC: 2.65 I: 2.02Serious
      Martino et al. (2017)
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      Pre -post intervention designICU (United states)AdultsSC: 6 months I: 24 monthsSC: 107 I: 153SC: 673 I: 1,272CVCCuros capCLABSISC: 7.4 I: 2.36Serious
      Merrill et al. (2014)
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      Pre -post intervention designICU and non-ICU (United states)Adults and childrenSC: 12 months I: 12 monthsSC: - I: -SC: 27,866 I: 26,489CVC, PICCCuros capCLABSISC: 1.5 I: 0.88Serious
      Milstone et al. (2021)
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      RCTNon-ICU (United states)ChildrenSC: 12 months I: 12 monthsSC: - I: -SC: 88,976 I: 88,421CVC, PICCCuros capCLABSISC: 1.27 I: 0.95Some concerns
      Pavia and Mazza (2016)
      • Pavia M
      • Mazza M
      Adding innovative practices and technology to central line bundle reduces bloodstream infection rate in challenging pediatric population.
      Pre -post intervention designNon-ICU (United states)ChildrenSC: 15 months I: 6 monthsSC: -I: 20-25SC: -I: -CVCSwabCapCLABSISC: 8.59 I: 3.89Critical
      Ramirez et al. (2012)
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      Pre -post intervention designICU (United states)AdultsSC: 12 months I: 12 monthsSC: -I: -SC: 2,105 I: 2,000CVC, PICC, portsCuros capCLABSISC: 1.9 I: 0.5Critical
      Rickard et al. (2021)
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      RCTNon-ICU (Australia)AdultsSC: 20 months I: 20 monthsSC: 61 I: 60SC: 725 I: 588CVCSwabCapCLABSISC: 1.38 I: 1.7Some concerns
      Stango et al. (2014)
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      Pre -post intervention designICU and non-ICU (United states)AdultsSC: 21 months I: 21 monthsSC: - I: -SC: 25,000 I: 22,892CVCSwabCapCLABSISC: 1.52 I: 0.83Critical
      Sweet et al. 2012
      • Sweet MA
      • Cumpston A
      • Briggs F
      • Craig M
      • Hamadani M
      Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line–associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit.
      Pre -post intervention designNon-ICU (United states)AdultsSC: 12 months I: 6 monthsSC: 472 I: 282SC: 6,851 I: 3,005CVC, PICC, implanted portCuros capCLABSISC: 2.3 I: 0.3Moderate
      Taşdelen Öğülmen et al. 2020
      • Tasdelen Ogulmen D
      • Ates S.
      Use of alcohol containing caps for preventing bloodstream infections: A randomized controlled trial.
      RCTICU (Turkey)AdultsSC: 5 months I: 5 monthsSC: 48 I: 47SC: 10.218 I: 8,460CVCCuros capCLABSISC: 1.37 I: 0.1Moderate
      Wright et al. (2013)
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      Pre -post intervention designICU and non-ICU (United states)AdultsSC: 9 months I: 18 monthsSC: 1,977 I: 2,860SC: 11,154 I: 18,972PICCSwabCapCLABSISC: 1.43 I: 0.69Serious
      Abbreviations: ‘-‘ = not mentioned; BSI= bloodstream infection; CLABSI = central line-associated bloodstream infection; CRBSI = catheter-related bloodstream infection; CVC = central venous catheter; ICU = intensive care unit; SC = standard care; PICC: peripherally inserted central catheter; PIV = peripheral intravenous catheter; RCT: randomised controlled trial; VAD: vascular access device; I= intervention (antiseptic barrier caps).
      1 CLABSI or CRBSI
      2 ROB2 or ROBINS-II

      Study characteristics

      Study characteristics are shown in Table 1. Ten studies evaluated the Curos cap, and six the SwabCap. Eleven studies included only adults, four exclusively children, and one study included both. Five studies exclusively focused on ICU patients, five on non-ICU, and six had a mixed population. All studies were conducted in the hospital, excluding Millstone et al. these authors conducted a RCT in pediatric hematology-oncology patients in the ambulatory setting.
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      We included three randomized controlled trials (RCTs) and twelve non-RCTs.

      Meta-analysis

      Figure 3 shows the individual and pooled data of all included studies. In total, 391 CLABSIs in 273,993 catheter days occurred in the intervention group with an incidence rate of 1.43/1,000 catheter days. In the standard care group, 620 CLABSIs in 284.912 catheter days occurred, with an incidence rate of 2.18/1,000 catheter days, resulting in a risk ratio of 0.65 (95%CI 0.55-0.76; P < 0.00001). Subgroup analyses are presented in Figures 4, 5, and 6. In eleven studies with adult patients, the risk for CLABSIs was reduced by 0.55 (95%CI 0.42-0.73; P < 0.0001) and in three studies with children a risk ratio of 0.75 (95%CI 0.58-0.96; P= 0.02) was found (Figure 4). Both type of brands (Curos and SwabCap) showed a decrease in CLABSI risk in favor of the ABCs of 0.59 (95%CI 0.46-0.77; P < 0.0001) and 0.71 (95%CI 0.59-0.85; P = 0.0002), respectively (Figure 5). Studies exclusively focusing on ICU populations showed a risk ratio of 0.36 (95%CI 0.15-0.88; P= 0.02). In the non-ICU population, a similar risk reduction was found, although non-significant (0.63, 95%CI 0.33-1.18; P=0.15) (Figure 6).
      Figure 3
      Figure 3Forrest plot all included studies - results are presented in two groups: antiseptic barrier caps (ABCs) versus standard care, with a risk ratio and 95% confidence intervals (95% CI).
      Figure 4
      Figure 4Forrest plot of subgroups (adults and children) - results are presented in two groups: antiseptic barrier caps (ABCs) versus standard care, with a risk ratio and 95% CI. Articles without separate outcomes on children and adults are not included.
      Figure 5
      Figure 5Forrest plot of subgroups (Curos cap and SwabCap) - results are presented in two groups: antiseptic barrier caps (ABCs) versus standard care, with a risk ratio and 95% CI.
      Figure 6
      Figure 6Forrest plot of subgroups (ICU and non-ICU) - results are presented in two groups: antiseptic barrier caps (ABCs) versus standard care, with a risk ratio and 95% CI. Articles without separate outcomes on intensive care unit (ICU) and non-ICU patients are not included.
      Heterogenicity between studies for the main outcome and almost all subgroup analyses were low, with an I2 of 0-40%. Only the ICU group showed a heterogenicity of 41%, which was considered substantial.

      Secondary outcomes

      Safety

      Only three studies reported safety data.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      ,
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      Two studies reported the use of Curos caps and did not observe damage to the access valves nor any other device-related adverse events.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      One study investigating SwabCap reported that two needleless connectors became opaque and additionally observed 70% IPA seeping between the rubber inner and outer plastic. However, this did not have any impact on the patient.
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.

      Costs

      Six studies evaluated costs of ABCs and reported significant annual cost savings.
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      ,
      • Kamboj M
      • Blair R
      • Bell N
      • Son C
      • Huang YT
      • Dowling M
      • et al.
      Use of Disinfection Cap to Reduce Central-Line-Associated Bloodstream Infection and Blood Culture Contamination Among Hematology-Oncology Patients.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      ,
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      ,
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      Calculated costs per 1,000 catheter days range from $10,000 to $39,000, with one outlier of $130,144 per 1,000 catheter days (Table 2). The average cost saving per 1,000 catheter days is $41,000.
      Table 2Calculated costs per 1,000 catheter days
      Annual costs savingsNumber of catheter daysCosts savings per 1,000 catheter days
      Cameron-Watson et al. (2016)
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      $740,5185,690$130,144
      Kamboj et al. (2015)
      • Kamboj M
      • Blair R
      • Bell N
      • Son C
      • Huang YT
      • Dowling M
      • et al.
      Use of Disinfection Cap to Reduce Central-Line-Associated Bloodstream Infection and Blood Culture Contamination Among Hematology-Oncology Patients.
      $3,268,99084,043$38,897
      Merrill et al. (2014)
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      $282,84084,043$10,407
      Ramirez et al. (2012)
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      $39,0502,053$19,021
      Stango et al. (2014)
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      $464,44023,346$19,395
      Wright et al. (2013)
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      $390,61715,063$25,932
      Cost savings per 1,000 catheter day are calculated from annual costs and number of catheter days. Annual costs were retrieved from the reported studies. The number of catheter days include mean catheter days of the standard care and ABC groups.

      Compliance and satisfaction

      Nine studies evaluated compliance when using ABCs
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      ,
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      ,
      • Sweet MA
      • Cumpston A
      • Briggs F
      • Craig M
      • Hamadani M
      Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line–associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit.
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      ,
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      , and three evaluated satisfaction.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      ,
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      ABCs were highly appreciated by health care workers due to time-savings and ease of use.
      • Sweet MA
      • Cumpston A
      • Briggs F
      • Craig M
      • Hamadani M
      Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line–associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit.
      ,
      • Wright MO
      • Tropp J
      • Schora DM
      • Dillon-Grant M
      • Peterson K
      • Boehm S
      • et al.
      Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
      The ease of use and simple monitoring due to the green color of the ABCs increased compliance.
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      All available studies reported a high satisfaction rate. A mean satisfaction of 9, on a scale of 0-10, was scored by 22 registered nurses.
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      In the PICU and the NICU, 28 nurses rated an overall mean of 9.2 and 8.6, respectively.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      Only one study evaluated ABC satisfaction in patients: 92% of the 1094 patients gave positive feedback and felt empowered by the ABCs.
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.

      Quality assessment and publication bias

      Overall risk of bias assessment is shown in Table 1. A complete assessment of each study is presented in Figure A1 and A2. In most studies, a substantial risk of bias was observed, mainly because of missing information, lack of correction for confounders, or simultaneous implementation of another intervention. A sensitivity analysis based on the degree of risk of bias (low to moderate, or high to critical) showed in both groups the same significant reduction in CLABSI risk in favor of ABCs, 0.66 (95%CI 0.49-0.90; P = 0.009) and 0.62 (95%CI 0.49-0.77; P < 0.00001), respectively (Figure A3). Sensitivity analyses based on the study design (RCT or non-RCT) showed in the non-RCT group a significant reduction in CLABSI risk 0.65 (95%CI 0.56-0.77; P < 0.00001), but not in the RCT group 0.46 (95%CI 0.11-1.91; P= 0.29) (Figure A4). Based on the asymmetric funnel plot presented in Figure 7, publication bias may be present.
      Figure 7
      Figure 7Funnel plot for assessing publication bias

      Discussion

      A safe and reliable central venous access is the cornerstone of treatments in numerous clinical settings. Despite various preventive measures, the presence of CVCs still poses a serious risk for developing septic complications. It should be emphasized that adequate CVC care with strict adherence to aseptic protocols when handling CVCs remains key to prevent CLABSIs. It is, however, widely recognized that strict compliance with these standard manual catheter care procedures remains difficult for many patients and even healthcare workers. Despite concerns with respect to the quality of the available studies, it seems that use of ABCs holds promise as a secondary measure to prevent CLABSIs, as overall analyses showed a protective effect of ABCs compared to standard care.
      Our analysis is a comprehensive update of previous meta-analyses, which showed a CLABSI rate reduction of 0.43-0.60, which is in line with the rate ratio of 0.65 currently found in this study.
      • Tejada S
      • Leal-Dos-Santos M
      • Peña-López Y
      • Blot S
      • Alp E
      • Rello J.
      Antiseptic barrier caps in central line-associated bloodstream infections: A systematic review and meta-analysis.
      ,
      • Voor In 't Holt AF
      • Helder OK
      • Vos MC
      • Schafthuizen L
      • Sülz S
      • van den Hoogen A
      • et al.
      Antiseptic barrier cap effective in reducing central line-associated bloodstream infections: A systematic review and meta-analysis.
      ,
      • Flynn JM
      • Larsen EN
      • Keogh S
      • Ullman AJ
      • Rickard CM.
      Methods for microbial needleless connector decontamination: A systematic review and meta-analysis.
      To date, only one small pilot study showed an increased infection rate with the use of ABCs, but, this concerned a non-significant trend most probably resulted from underpowering.
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      Taken together, the overall trend is in favor of the ABCs, despite the low quality of the majority of the articles.
      In this study we performed several subgroup analyses. Almost all groups showed a clear decrease in CLABSI rates with the use of ABCs. Even in younger patients, who tend to be less compliant when managing CVADs due distractions or a more active lifestyle, fewer CLABSIs were seen when using ABCs (Figure 4). This protective effect of ABCs even applied to the ICU population, where CVCs are more often daily accessed, normally resulting in increased risks for CLABSIs.
      • Marschall J
      • Mermel LA
      • Fakih M
      • Hadaway L
      • Kallen A
      • O'Grady NP
      • et al.
      Strategies to Prevent Central Line-Associated Bloodstream Infections in Acute Care Hospitals: 2014 Update.
      These results show that a simple intervention, screwing an ABC on a CVC connector, prevents various risk full moments during catheter care and illustrates a strong universal effect within different patient populations. That such an effect was not seen in the non-ICU population is probably related due to the relatively small number of studies included, as the (non-significant) trend towards a protective effect for ABCs was similar as in the other subgroups. As head-to-head comparisons between ABC brands (Curos caps en SwabCaps) are currently lacking, it is difficult to establish which brand performs better. Subgroup analyses suggest similar reductions in CLABSI rates for Curos caps and SwabCaps, which may be explained by the close resemblance of both caps: both contain a 70% IPA impregnated sponge that passively disinfects the outside surface of the needle-less connectors. There are only minor differences in cap design, for example in the flexibility of the IPA impregnated sponge.
      • Voor In 't Holt AF
      • Helder OK
      • Vos MC
      • Schafthuizen L
      • Sülz S
      • van den Hoogen A
      • et al.
      Antiseptic barrier cap effective in reducing central line-associated bloodstream infections: A systematic review and meta-analysis.
      The sensitivity analysis of the groups ‘low to moderate risk of bias’, ‘high to critical risk of bias’ and ‘non-RCT’ all showed a significant decrease in CLABSI in favor of the caps, in contrast to the RCT group which showed a non-significant decrease in CLABSIs. This can partially be explained by the (underpowered) pilot study of Rickard et al.
      • Rickard CM
      • Flynn J
      • Larsen E
      • Mihala G
      • Playford EG
      • Shaw J
      • et al.
      Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
      Furthermore, one of the three RCTs was the only conducted study in the ambulatory setting and focused on the pediatric population.
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      Therefore, more research should show the effect of the ABC in the ambulatory setting.
      Remarkedly, only a few studies report safety aspects concerning the use of ABCs. Sauron et al. raised concerns about the safety of 70% IPA impregnated ABCs as several access valves malfunctioned shortly after using SwabCaps in their neonatal ICU. The authors investigated IPA concentrations in the bloodstream after infusate passed through the catheter. They found that IPA leaked from the SwabCap through the access valve into the bloodstream, which resulted in IPA concentrations exceeding critical limits for premature neonates. However, the risk for IPA intoxication should be minimized if the needleless connector is dried in open air before drugs infusion.
      • Sauron C
      • Jouvet P
      • Pinard G
      • Goudreault D
      • Martin B
      • Rival B
      • et al.
      Using isopropyl alcohol impregnated disinfection caps in the neonatal intensive care unit can cause isopropyl alcohol toxicity.
      Sauron et al. also assessed whether SwabCaps affected the function and appearance of catheter access valves. While structural macroscopic changes were noted in more than half of the tested valves, none of them malfunctioned.
      • Sauron C
      • Jouvet P
      • Pinard G
      • Goudreault D
      • Martin B
      • Rival B
      • et al.
      Using isopropyl alcohol impregnated disinfection caps in the neonatal intensive care unit can cause isopropyl alcohol toxicity.
      The importance of compliance to use ABCs was stressed by Merrill et al., who found a strong association between the use of Curos caps and a decrease in CLABSI rate: a 10% increase in compliance resulted in a 7% drop in CLABSI rate (incidence rate ratio 0.93).
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      This is in line with other studies stating that manual disinfection alone is not sufficient to prevent CLABSIs and compliance is an important factor to improve CLABSI prevention.

      Vries d. Reducing Bloodstream Infection Risk in Central and Peripheral Intravenous Lines: Initial Data on Passive Intravenous Connector Disinfection. 2014.

      ,
      • Pavia M
      • Mazza M
      Adding innovative practices and technology to central line bundle reduces bloodstream infection rate in challenging pediatric population.
      ,
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      ,
      • Menyhay SZ
      • Maki DG.
      Disinfection of needleless catheter connectors and access ports with alcohol may not prevent microbial entry: the promise of a novel antiseptic-barrier cap.
      All studies evaluating compliance of ABCs emphasize the ease of use and time-saving aspect of ABCs over standard rubbing methods.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      ,
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      In addition, the bright color of ABCs was a valued by caregivers as it highlights the catheter access point and makes compliance easy to monitor.
      • Helder OK
      • van Rosmalen J
      • van Dalen A
      • Schafthuizen L
      • Vos MC
      • Flint RB
      • et al.
      Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
      ,
      • Martino A
      • Thompson L
      • Mitchell C
      • Trichel R
      • Chappell W
      • Miller J
      • et al.
      Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
      When ABCs are placed on a CVC, unlike with a manual procedure, disinfection of the needleless connector is visualized for all caregivers. Despite the advantages of ABCs over standard manual disinfection procedures alone, compliance is not always guaranteed. Stango et al. found that healthcare workers needed time to adjust to the implementation of SwabCaps and that compliance increased when the use of ABCs was monitored (when observed in place on the connector) and the caps were easily available.
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      The latter was exemplified by a study of Ramirez et al., who noted a broad range in compliance (25-100%), but in particular a low compliance related to poor cap availability in the patient's room. After cap strips were hung on intravenous poles at the bedside, compliance increased and resulted in an average compliance of 73% throughout their trial.
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      It is therefore important to note that education of staff members, monitoring of ABC use, and availability of caps is key to improve and maintain compliance.
      • Stango C
      • Runyan D
      • Stern J
      • Macri I
      • Vacca M.
      A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
      ,
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      ,
      • Inchingolo R
      • Pasciuto G
      • Magnini D
      • Cavalletti M
      • Scoppettuolo G
      • Montemurro G
      • et al.
      Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit 11 Medical and Health Sciences 1117 Public Health and Health Services 11 Medical and Health Sciences 1103 Clinical Sciences.
      CLABSIs have a high impact on patient's experienced quality of life, hospital admission length, and subsequently costs. The Centers for Disease Control and Prevention estimate that for every CLABSI, hospital stay increases by two to five days.
      • O'Grady NP
      • Alexander M
      • Burns LA
      • Dellinger EP
      • Garland J
      • Heard SO
      • et al.
      Guidelines for the prevention of intravascular catheter-related infections.
      Cost-effectiveness is an important issue when implementing any device. Although costs of a single cap are low (around $0.25-0.50), each use of the CVC requires a new cap, and in some patients, this implies the use of several caps per day. Cost effectiveness is also related to CLABSI rates, which differ significantly between patient populations (e.g., pediatric versus adults). All included studies in this review concluded that implementation of ABCs not only decreased CLABSI rates, but also associated healthcare costs. However, interpretation remains difficult since all authors reported absolute costs, which vary per study population. In addition, most studies only calculated product- and CLABSI costs but did not perform an extensive cost-effect analysis. Therefore, these results should be interpreted with caution.

      Strengths and limitations

      This study has several strengths. First, we performed a comprehensive literature search to identify all available latest prospective studies. If studies lacked outcome data, we actively contacted corresponding authors for more information. Eventually, this resulted in an inclusion of 6 extra (both old and new) studies, including two new RCTs, which were not reported in the latest meta-analysis.
      • Ramirez C
      • Lee AM
      • Welch K.
      Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
      ,
      • Merrill KC
      • Sumner S
      • Linford L
      • Taylor C
      • Macintosh C.
      Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
      ,
      • Cameron-Watson C.
      Port protectors in clinical practice: an audit.
      • Castello FV
      • Maher A
      • Cable G.
      Reducing bloodstream infections in pediatric rehabilitation patients receiving parenteral nutrition.
      • Cruz-Aguilar R
      • Carney J
      • Mondaini V
      • Vehreschild M
      • Griskaitis M
      • Salmanton-García J
      • et al.
      A quality improvement study on the reduction of central venous catheter-associated bloodstream infections by use of self-disinfecting venous access caps (STERILE).
      • Milstone AM
      • Rosenberg C
      • Yenokyan G
      • Koontz DW
      • Miller MR.
      Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
      Second, previous meta-analyses were hampered by inclusion of only one single pediatric study, which precluded subgroup analyses of these patients, whereas we were able to track two additional studies. Finally, the data that we obtained allowed analyses on catheter level instead of patient level, which, in addition to the sensitivity analyses, results in more precise and reliable outcomes.
      As with all reviews, the quality of the present study mainly builds on the individual quality of included studies. Heterogenicity between studies and their (mostly) high risk of bias remains a limitation that may have impacted our meta-analyses. The majority of studies concerned pre- and post-intervention designs, and only three RCTs were available. In addition, the variety of (sometimes mixed) patients populations precluded several subgroup analyses, such as those for oncology or hematology patients. On the other hand, while the heterogenicity is substantial (which is inherently associated with the broad range in clinical patient populations that require a CVC), the overall message is that all studies provided support for the use of ABCs. Last, relevant information on infection prevention, such as the use of antimicrobial locks, is missing in some manuscripts and therefore it is not always clear whether the effect of ABCs is on top of other antiseptic measures.
      In conclusion, the most recent clinical evidence keeps providing support for the use of ABCs for passive and continuous disinfection of needle-less connectors, without a need for the more cumbersome manual needleless connector disinfection. Despite the limited methodological quality of most available studies, the overall verdict at this point seems to be that ABCs establish an effective strategy to reduce CLABSI rates in patient populations that depend on a CVC. These caps are safe, highly appreciated by health care workers for their ease of use, are timesaving in clinical practice and there are no disadvantages with their use. High-quality and controlled trials in various patient populations with a different infection risk remain necessary to unequivocally establish that ABCs are cost-effective.

      Author contributions

      VG: Conceptualization, methodology, validation, formal analysis, investigation, data curation, writing - original draft, writing - review and editing, visualization, project administration
      MvE: Conceptualization, methodology, validation, investigation, data curation, writing - original draft, writing - review and editing
      YW: Conceptualization, methodology, validation, writing - review and editing
      GW: Conceptualization, methodology, validation, writing - review and editing, supervision

      Conference presentation

      None

      Funding

      This research did not receive any funding

      Data sharing

      Data described in the manuscript, code book, and analytic code will be made available upon request pending approval

      CRediT authorship contribution statement

      Veerle ELM Gillis: Conceptualization, Methodology, Validation, Formal analysis, Investigation, Data curation, Writing – original draft, Writing – review & editing, Visualization, Project administration. Marijn J van Es: Conceptualization, Methodology, Validation, Investigation, Data curation, Writing – original draft, Writing – review & editing. Yannick Wouters: Conceptualization, Methodology, Validation, Writing – review & editing. Geert JA Wanten: Conceptualization, Methodology, Validation, Writing – review & editing, Supervision.

      Conflicts of Interest and financial disclosure

      VG: None
      MvE: None
      YW: None
      GW: reports grants from Geistlich Pharma, and consulting fees from Geistlich Pharma and Zealand outside the submitted work.

      Acknowledgements

      We would like to thank the authors of the included publications who provided extra information

      Appendix. SUPPLEMENTARY MATERIALS

      References

        • Maki DG
        • Kluger DM
        • Crnich CJ.
        The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies.
        Mayo Clin Proc. 2006; 81: 1159-1171
        • Ruesch S
        • Walder B
        • Tramèr MR.
        Complications of central venous catheters: internal jugular versus subclavian access–a systematic review.
        Crit Care Med. 2002; 30: 454-460
        • Pronovost P
        • Needham D
        • Berenholtz S
        • Sinopoli D
        • Chu H
        • Cosgrove S
        • et al.
        An intervention to decrease catheter-related bloodstream infections in the ICU.
        N Engl J Med. 2006; 355: 2725-2732
        • Ziegler MJ
        • Pellegrini DC
        • Safdar N.
        Attributable mortality of central line associated bloodstream infection: systematic review and meta-analysis.
        Infection. 2015; 43: 29-36
        • Helder OK
        • van Rosmalen J
        • van Dalen A
        • Schafthuizen L
        • Vos MC
        • Flint RB
        • et al.
        Effect of the use of an antiseptic barrier cap on the rates of central line–associated bloodstream infections in neonatal and pediatric intensive care.
        American Journal of Infection Control. 2020; 48: 1171-1178
        • de Cicco M
        • Panarello G
        • Chiaradia V
        • Fracasso A
        • Veronesi A
        • Testa V
        • et al.
        Source and route of microbial colonisation of parenteral nutrition catheters.
        Lancet. 1989; 2: 1258-1261
        • Frasca D
        • Dahyot-Fizelier C
        • Mimoz O.
        Prevention of central venous catheter-related infection in the intensive care unit.
        Crit Care. 2010; 14: 212
        • Bond A
        • Chadwick P
        • Smith TR
        • Nightingale JMD
        • Lal S.
        Diagnosis and management of catheter-related bloodstream infections in patients on home parenteral nutrition.
        Frontline Gastroenterol. 2020; 11: 48-54
        • Mermel LA
        • Allon M
        • Bouza E
        • Craven DE
        • Flynn P
        • O'Grady NP
        • et al.
        Clinical practice guidelines for the diagnosis and management of intravascular catheter-related infection: 2009 Update by the Infectious Diseases Society of America.
        Clinical infectious diseases: an official publication of the Infectious Diseases Society of America. 2009; 49: 1-45
        • Casey AL
        • Karpanen TJ
        • Nightingale P
        • Elliott TSJ.
        An in vitro comparison of standard cleaning to a continuous passive disinfection cap for the decontamination of needle-free connectors.
        Antimicrob Resist Infect Control. 2018; 7: 50
        • Stango C
        • Runyan D
        • Stern J
        • Macri I
        • Vacca M.
        A successful approach to reducing bloodstream infections based on a disinfection device for intravenous needleless connector hubs.
        J Infus Nurs. 2014; 37: 462-465
      1. Vries d. Reducing Bloodstream Infection Risk in Central and Peripheral Intravenous Lines: Initial Data on Passive Intravenous Connector Disinfection. 2014.

        • Pavia M
        • Mazza M
        Adding innovative practices and technology to central line bundle reduces bloodstream infection rate in challenging pediatric population.
        American journal of infection control. 2016; 44: 112-114
        • Ramirez C
        • Lee AM
        • Welch K.
        Central venous catheter protective connector caps reduce intraluminal catheter-related infection.
        JAVA - Journal of the Association for Vascular Access. 2012; 17: 210-213
        • Inchingolo R
        • Pasciuto G
        • Magnini D
        • Cavalletti M
        • Scoppettuolo G
        • Montemurro G
        • et al.
        Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit 11 Medical and Health Sciences 1117 Public Health and Health Services 11 Medical and Health Sciences 1103 Clinical Sciences.
        BMC infectious diseases. 2019; : 19
        • Kamboj M
        • Blair R
        • Bell N
        • Son C
        • Huang YT
        • Dowling M
        • et al.
        Use of Disinfection Cap to Reduce Central-Line-Associated Bloodstream Infection and Blood Culture Contamination Among Hematology-Oncology Patients.
        Infect Control Hosp Epidemiol. 2015; 36: 1401-1408
        • Martino A
        • Thompson L
        • Mitchell C
        • Trichel R
        • Chappell W
        • Miller J
        • et al.
        Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU.
        Burns. 2017; 43: 956-964
        • Merrill KC
        • Sumner S
        • Linford L
        • Taylor C
        • Macintosh C.
        Impact of universal disinfectant cap implementation on central line-associated bloodstream infections.
        Am J Infect Control. 2014; 42: 1274-1277
        • Tejada S
        • Leal-Dos-Santos M
        • Peña-López Y
        • Blot S
        • Alp E
        • Rello J.
        Antiseptic barrier caps in central line-associated bloodstream infections: A systematic review and meta-analysis.
        Eur J Intern Med. 2022;
        • Sterne JA
        • Hernán MA
        • Reeves BC
        • Savović J
        • Berkman ND
        • Viswanathan M
        • et al.
        ROBINS-I: a tool for assessing risk of bias in non-randomised studies of interventions.
        BMJ (Clinical research ed). 2016; 355 (-i): i4919
        • Sterne JAC
        • Savović J
        • Page MJ
        • Elbers RG
        • Blencowe NS
        • Boutron I
        • et al.
        RoB 2: a revised tool for assessing risk of bias in randomised trials.
        BMJ. 2019; 366: l4898
        • McGuinness LA
        • Higgins JPT.
        Risk-of-bias VISualization (robvis): An R package and Shiny web app for visualizing risk-of-bias assessments.
        Research Synthesis Methods. 2020; (n/a)
        • Higgins JP
        • Thompson SG
        • Deeks JJ
        • Altman DG.
        Measuring inconsistency in meta-analyses.
        Bmj. 2003; 327: 557-560
        • Borenstein M
        • Higgins JP.
        Meta-analysis and subgroups.
        Prev Sci. 2013; 14: 134-143
        • Tasdelen Ogulmen D
        • Ates S.
        Use of alcohol containing caps for preventing bloodstream infections: A randomized controlled trial.
        Journal of Vascular Access. 2020;
        • Voor In 't Holt AF
        • Helder OK
        • Vos MC
        • Schafthuizen L
        • Sülz S
        • van den Hoogen A
        • et al.
        Antiseptic barrier cap effective in reducing central line-associated bloodstream infections: A systematic review and meta-analysis.
        Int J Nurs Stud. 2017; 69: 34-40
        • Sweet MA
        • Cumpston A
        • Briggs F
        • Craig M
        • Hamadani M
        Impact of alcohol-impregnated port protectors and needleless neutral pressure connectors on central line–associated bloodstream infections and contamination of blood cultures in an inpatient oncology unit.
        American Journal of Infection Control. 2012; 40: 931-934
        • Wright MO
        • Tropp J
        • Schora DM
        • Dillon-Grant M
        • Peterson K
        • Boehm S
        • et al.
        Continuous passive disinfection of catheter hubs prevents contamination and bloodstream infection.
        Am J Infect Control. 2013; 41: 33-38
        • Cameron-Watson C.
        Port protectors in clinical practice: an audit.
        Br J Nurs. 2016; 25: S25-S31
        • Castello FV
        • Maher A
        • Cable G.
        Reducing bloodstream infections in pediatric rehabilitation patients receiving parenteral nutrition.
        Pediatrics. 2011; 128: e1273-e1278
        • Cruz-Aguilar R
        • Carney J
        • Mondaini V
        • Vehreschild M
        • Griskaitis M
        • Salmanton-García J
        • et al.
        A quality improvement study on the reduction of central venous catheter-associated bloodstream infections by use of self-disinfecting venous access caps (STERILE).
        Am J Infect Control. 2021; 49: 586-592
        • Milstone AM
        • Rosenberg C
        • Yenokyan G
        • Koontz DW
        • Miller MR.
        Alcohol-impregnated caps and ambulatory central-line-associated bloodstream infections (CLABSIs): A randomized clinical trial.
        Infect Control Hosp Epidemiol. 2021; 42: 431-439
        • Rickard CM
        • Flynn J
        • Larsen E
        • Mihala G
        • Playford EG
        • Shaw J
        • et al.
        Needleless connector decontamination for prevention of central venous access device infection: a pilot randomized controlled trial.
        American journal of infection control. 2021; 49: 269-273
        • Page MJ
        • McKenzie JE
        • Bossuyt PM
        • Boutron I
        • Hoffmann TC
        • Mulrow CD
        • et al.
        The PRISMA 2020 statement: an updated guideline for reporting systematic reviews.
        BMJ. 2021; 372: n71
        • Flynn JM
        • Larsen EN
        • Keogh S
        • Ullman AJ
        • Rickard CM.
        Methods for microbial needleless connector decontamination: A systematic review and meta-analysis.
        Am J Infect Control. 2019; 47: 956-962
        • Marschall J
        • Mermel LA
        • Fakih M
        • Hadaway L
        • Kallen A
        • O'Grady NP
        • et al.
        Strategies to Prevent Central Line-Associated Bloodstream Infections in Acute Care Hospitals: 2014 Update.
        Infection Control & Hospital Epidemiology. 2014; 35: S89-S107
        • Sauron C
        • Jouvet P
        • Pinard G
        • Goudreault D
        • Martin B
        • Rival B
        • et al.
        Using isopropyl alcohol impregnated disinfection caps in the neonatal intensive care unit can cause isopropyl alcohol toxicity.
        Acta Paediatr. 2015; 104: e489-e493
        • Menyhay SZ
        • Maki DG.
        Disinfection of needleless catheter connectors and access ports with alcohol may not prevent microbial entry: the promise of a novel antiseptic-barrier cap.
        Infect Control Hosp Epidemiol. 2006; 27: 23-27
        • O'Grady NP
        • Alexander M
        • Burns LA
        • Dellinger EP
        • Garland J
        • Heard SO
        • et al.
        Guidelines for the prevention of intravascular catheter-related infections.
        Clin Infect Dis. 2011; 52: e162-e193